| Model Number |
EEG-1200A |
| Medical Device Problem Codes |
Overheating of Device (1437); Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The monitor technician reported that there was a burning smell from the system.They did some troubleshooting and found that the power supply inside the computer was causing the smell.They replaced the part with the nihon kohden provided part number for the replacement power supply.Issue was resolved by the customer.This was not in patient use and no other devices being used with the system.
|
| |
|
Event or Problem Description
|
|
The monitor technician reported that there was a burning smell from the system.They did some troubleshooting and found that the power supply inside the computer was causing the smell.They replaced the part with the nihon kohden provided part number for the replacement power supply.Issue was resolved by the customer.This was not in patient use.
|
| |
|