| Model Number |
N/A |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); Insufficient Information (4580)
|
| Date of Event |
07/12/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medical products: proximal humerus, left, long, 8.5x220mm; catalog#: 47-2496-221-08; lot#: 3031559.Blunt tip screw, 4x36mm; catalog#: 47-2486-036-40; lot#: 3010630.Blunt tip screw, 4x38mm; catalog#: 47-2486-038-40; lot#: 3024680.Blunt tip screw, 4x38mm; catalog#: 47-2486-038-40; lot#: 3064861.Cortical bone screw, 4x20mm; catalog#: 47-2486-120-40; lot#: 2999909.Cortical bone screw, 4x22mm; catalog#: 47-2486-122-40; lot#: 3006524.Cortical bone screw, 4x26mm; catalog#: 47-2486-126-40; lot#: 3024356.Proximal humerus nail cap, 0mm; catalog#: 47-2488-010-00; lot#: 3036115.Therapy date: (b)(6) 2021.The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.Where lot numbers were received for the devices, the device history records were reviewed and found to be conforming.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).
|
| |
|
Event or Problem Description
|
|
Patient was implanted with ann nail on an unknown side and one month post-implantation it was noticed that one of the proximal screws had backed out from its implanted position.Hence, patient underwent a revision surgery during which the backed out screw was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being filled to relay additional information, which was unknown at the time of the previous medwatch.D10: medical product: torque limiting handle; catalog#: 27923; lot#: unknown.Lg cann screwdriver handle; catalog#: 214149000; lot#: unknown.Therapy date: (b)(6) 2021 additional information was received on jul 21, 2021.The manufacturer received other source documents for review.Should additional information become available and / or the device(s) be returned for evaluation and an investigation result be available, that changes this assessment, an amended medical device report will be submitted.Zimmer¿s reference number of this file is (b)(4).
|
| |
|
Event or Problem Description
|
|
Patient was implanted on an unknown side.1 month post the surgery, it was noticed that the proximal screws gradually migrated from the implanted position.Patient underwent revision surgery during which the migrated screw was explanted.The torque driver was used to engage the corelock, furthermore non-torque driver was used to tighten more after using the torque driver.
|
| |
|
Event or Problem Description
|
|
No event update.Investigation results are now available.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of event description: it was reported that the patient was implanted on (b)(6) 2021.1 month post the surgery, it was noticed that the proximal screws gradually migrated from the implanted position.A revision surgery has been performed on (b)(6) 2021, during which, only the migrated screw was explanted.Patent had a fall during rehabilitation.Review of received data: - due diligence: further due diligence to support the conclusion was completed and documented in diligence log.Product evaluation: - no product was returned; therefore, visual and dimensional evaluation could not be performed.Review of product documentation: - device purpose: this device is intended for treatment.- product compatibility: the product combination was approved by zimmer biomet.- dhr review: review of the device history records identified no deviations or anomalies during manufacturing.- raw material certificate: the raw material certificate was reviewed with no anomalies noted.- surgical technique sap: the surgical technique 197-glbl-en explains that the locking of the corelock is done using the corelock driver with torque limiting handle.Turn slowly clockwise to tighten and engage the corelock mechanism until a click is felt from the torque limiting handle.Conclusion: it was reported that the patient was implanted on (b)(6) 2021.1 month post the surgery, it was noticed that the proximal screws gradually migrated from the implanted position.A revision surgery has been performed on (b)(6) 2021, during which, only the migrated screw was explanted.Patent had a fall during rehabilitation.Neither x-rays, operative notes, office visit notes, nor device(s) or photos of the device(s) were received; therefore the condition of the component(s) is unknown.It is reported that the patient fell during rehabilitation.Nevertheless, it remains unknown what potential effect the fall had to the reported event.The locking of the corelock during the initial surgery has not been performed as specified in the surgical technique using only the corelock driver with torque limiting handle.Instead, additionally a non-torque limiting screwdriver was used.It remains unknown what the potential effect of this deviation from the surgical technique could be.Based on the investigation the reported event cannot be confirmed.The quality records show that all specified characteristics have met the specifications valid at the time of production.Therefore, the investigation did not identify a nonconformance or a complaint out of box (coob).Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition and behavior, implantation procedure or design features.If and to what extent any of these aspects may have influenced the backing out of the screw remains unknown.The need for corrective measures is not indicated for the time being and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet's reference number of this file is (b)(4).
|
| |
|