Manufacturing review: the device history record review could not be performed as the lot number is unknown.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately five months of post deployment, the computed tomography images displayed the inferior vena cava filter with the tip at the level of the renal veins.There is a 13-millimeter hypoattenuating filling defect just above the inferior vena cava filter and may represent focal nonocclusive thrombus.Therefore, the investigation is inconclusive as no objective evidence has been provided to confirm any alleged deficiency with the filter.Additionally, it can be confirmed that the patient experienced thrombus above the filter post deployment.However, the relationship to the filter is unknown.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
|