• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0576-01
Device Problems Off-Label Use (1494); Calibration Problem (2890); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the customer was unable to calibrate and irrational blood pressure readings were being displayed.The insertion was reported to be axillary, which is not the method described in the device instructions for use.The balloon is functioning perfectly, but has the classic different arterial waveforms and issues with calibration common to axillary insertions.There was no patient harm or adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period (b)(6) to (b)(6) was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 8FR. 50CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
MDR Report Key12225116
MDR Text Key263404423
Report Number2248146-2021-00478
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108605
UDI-Public10607567108605
Combination Product (y/n)N
PMA/PMN Number
K112327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0684-00-0576-01
Device Catalogue Number0684-00-0575
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/06/2021
Initial Date FDA Received07/26/2021
Supplement Dates Manufacturer Received09/22/2021
Supplement Dates FDA Received09/22/2021
Patient Sequence Number1
-
-