| Model Number |
1116156 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Chest Pain (1776)
|
| Date of Event |
07/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a bipap a40 device's sound abatement foam became degraded and caused severe chest pain.The patient was admitted to the hospital for two weeks.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported mdr 2518422-2021-02475 with the incorrect investigation conclusion code of code 12: cause traced to device design, in section h6.The correct investigation conclusion code is code 11: conclusion not yet available.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported an allegation of an issue related to sound abatement foam.This action was reported to fda per 21 cfr part 806.The device will be corrected per res 88058.Corrected data: d8 was this device serviced by a third party?: previously reported as no ; updated to unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging a bipap a40 device's sound abatement foam became degraded and caused severe chest pain.The patient was admitted to the hospital for two weeks.The device was returned to the manufacturer's product investigation laboratory for investigation.During investigation visually inspected the a40 silver series and did not observe any visual defects.Reviewed error logs and consulted engineering.Nothing abnormal appeared, installed a bacteria filter on the outlet of the unit and ran the unit at the received settings, on an asl lung for an hour at approximately 500ml volume without error.Removed the bacteria filter and inspected it.No foam was observed in the filter.Unit operates as expected.The apnea alarm and the circuit disconnect alarm were checked.Both alarms operated as expected.
|
| |
|