| Model Number |
EPK-I5500C |
| Medical Device Problem Code |
Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary : it was caused due to the part failure in the processor.This device is classified as import for export, therefore 510k is not applicable.Model epk-i5500c-us is available in the usa with a 510k number epk-i5500c-us.This report is being filed as part of the pentax backlog management plan.
|
| |
|
Event or Problem Description
|
|
Imagina processor "freezing".The time of event is unknown.There was no report of patient harm.Date of event not known for the exact date, we just know the year the complaint was sent in to manufacturer.
|
| |
|
Search Alerts/Recalls
|