Model Number NAV2132
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding patient implanted with a spinal product using an inserter for a l5-s1 tlif extension and revision of hardware at l3-l5.It was reported that the side metal pieces of the elevate inserter that hug the elevate cage broke off.According to the physician, both pieces were extracted.The instrument broke and there was no fragment of the instrument remaining in the patient's body.There were no patient symptoms or complications as a result of the event.The product was used in operation field and came in contact with the patient.The pre-op diagnosis was lumbar instability neurogenic claudication due to lumbar spinal stenosis and levels implanted were l5-s1.There was no treatment or additional surgery performed as a result of this event.The product was replaced by a medtronic product.The patient's medical history included diabetes, blood pressure, arthritis.No further complications were reported/ anticipated.
 
Additional Manufacturer Narrative
D4: updated udi# h3: product analysis part# nav2132, lot# dn19h001- visual and optical inspection confirmed that both of the prongs at the tip of the inserter have been broken off.This type of damage is consistent with bend stress overload.H6: updated eval.Code method and eval.Code result post analysis medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATE¿ SPINAL SYSTEM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12230511
Report Number1030489-2021-00973
Device Sequence Number3036410
Product Code OLO
UDI-Device Identifier00643169771000
UDI-Public00643169771000
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K163581
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 11/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNAV2132
Device Catalogue NumberNAV2132
Device Lot NumberDN19H001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2021
Supplement Date Received by Manufacturer11/09/2021
Initial Report FDA Received Date07/27/2021
Supplement Report FDA Received Date11/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight94 KG
-
-