| Model Number |
25840018580 |
| Medical Device Problem Codes |
Break (1069); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.Information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The information was received from healthcare provider via a manufacturing representative regarding a patient with scoliosis at t10- pelvis for spinal therapy.It was reported that set screws popped off or let the rod slip out.There was a revision surgery that was required to place new implants that lasted 4 hours as a result of this event.It was unknown which screw popped off and surgeon removed all the screw in pile together.There was no further information reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: part # 25840018580; lot # ca18k077 visual and optical inspection did not reveal any damage to the head or the threads of the screw.The saddle has some wear on it from the seating and tightening of the rod.Functional check with a sample break off screw confirmed the screw would thread into the head without any issues.Unable to determine root cause of screw backing out.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|