Model Number DTMA1Q1
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 07/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: 694758 lead implanted (b)(6) 2007.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that there was a suspected inappropriate shock for atrial fibrillation (af) with rapid ventricular response that the cardiac resynchronization therapy defibrillator (crt-d) classified as ventricular fibrillation (vf).The device remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Follow-up concluded that the device system was functioning normally and the delivered shocks were deemed to be appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI QUAD CRT-D SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key12232430
Report Number3004209178-2021-11465
Device Sequence Number14460179
Product Code NIK
UDI-Device Identifier00643169705760
UDI-Public00643169705760
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/14/2018
Device Model NumberDTMA1Q1
Device Catalogue NumberDTMA1Q1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/06/2021
Supplement Date Received by Manufacturer08/12/2021
Initial Report FDA Received Date07/27/2021
Supplement Report FDA Received Date08/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
459888 LEAD, 407645 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age71 YR
Patient Weight77
-
-