| Model Number |
2000 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that guidewire tip detachment occurred.A 182cm j choice guidewire was selected for use; however, upon removal from the protective case, it was noticed that the guidewire tip was damaged and broken.The procedure was completed with another of same device.No complications were reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: unit returned with its original box labeled batch 26480125.The unit returned has tip damaged, as part of overall visual revision.The returned guidewire had the spring tip bent/kink.The spring tip is stretched.The device was measured in three sections (distal - medium - proximal) along it in order to confirm that is within specification.
|
| |
|
Event or Problem Description
|
|
It was reported that guidewire tip detachment occurred.A 182cm j choice guidewire was selected for use; however, upon removal from the protective case, it was noticed that the guidewire tip was damaged and broken.The procedure was completed with another of same device.No complications were reported and the patient status was stable.
|
| |
|
Search Alerts/Recalls
|