Model Number 3186
Device Problem High impedance (1291)
Patient Problems Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2021
Event Type  Injury  
Manufacturer Narrative
Date of event estimated.
 
Event Description
It was reported one lead has high impedance preventing the patient from entering mri mode.In turn, surgical intervention may take place in the future.Both of the patient's leads are being reported because it's unknown which one of the leads has high impedance.
 
Event Description
It was reported that the patient's lead was explanted and replaced.Surgical intervention resolved the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
PERCUTANEOUS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key12233308
MDR Text Key263831053
Report Number3006705815-2021-03656
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/25/2020
Device Model Number3186
Device Catalogue Number3186
Device Lot NumberA000050928
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/07/2021
Initial Date FDA Received07/27/2021
Supplement Dates Manufacturer Received09/01/2021
Supplement Dates FDA Received09/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
3186 SCS LEAD; 3186 SCS LEAD
Patient Weight70
-
-