| Model Number |
72290118 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the during hip arthroscopy - labrum repair, inside the patient, the 1.8mm q-fix disposable flexible drill broke off.All pieces were removed.The procedure was finished with a smith and nephew backup device.Surgical delay less than or equal to 30 minutes.Patient injuries were not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found the tip of the drill is broken.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.The complaint was confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device tip, or an inadvertent impact event.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|