ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 130, RIGHT, 11.5 MM, 21.5 CM; ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products item# 47248510510, lot# 3026107, znn, cmn lag screw, 10.5 mm, 105 mm including set screw.Item# 47248403550, lot# 64850532, 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head.Event description: it was reported that the patient was implanted on an unknown date and underwent a revision surgery on (b)(6) 2021 due to implant fracture.Review of received data: due diligence: no further due diligence required as all required information to support the conclusion is available or was already requested.X-rays: x-ray was reviewed by a radiologist.There is a comminuted femoral neck fracture with femoral nail and screw fixation.The femoral nail is fractured at the screw level with varus angulation at the nail fracture site.The screw is intact.The distal aspect of the nail is not included on the image.Images: review of the received images shows that the nail was broken through the hole for the lag screw.Based on the images the nature of the fracture cannot be evaluated.Surgical report: there is an implant surgery sheet and an explant surgery sheet available with some information like product ref and lot numbers.No other relevant information is available.Product evaluation: no product was returned; therefore, visual and dimensional evaluation could not be performed.Review of product documentation: device purpose: all involved devices are intended for treatment.Product compatibility: the compatibility check was performed from www.Productcompatibility.Zimmer.Com and showed that the product combination was approved by zimmer biomet.Dhr review: review of the device history records identified no deviations or anomalies during manufacturing.Raw material certificate: the raw material certificate was reviewed with no anomalies noted.Conclusion: it was reported that the patient was implanted on an unknown date and underwent a revision surgery on (b)(6) 2021 due to implant fracture.The quality records show that all specified characteristics (material, dimensions, surface, etc.) for the znn nail have met the specifications valid at the time of production.No product was returned; therefore, visual and dimensional evaluation could not be performed.The implant date of the znn nailing system is unknown.Therefore, it cannot be said how long the nail was in vivo.The removal was on (b)(6) 2021.Review of the received images shows that the nail was broken through the hole for the lag screw.Review of the received images shows that the nail was broken through the hole for the lag screw.However, based on the images the nature of the fracture cannot be evaluated.It is possible that the nail was fractured due to fatigue which can occur due to a cyclic overloading.Possible contributing factors to the overload could be a not properly healed bone fracture and / or not adherence to the postoperative protocol (patient behavior).However, based on the available information and performed investigation, an exact root cause for the nail breakage could not be determined.The investigation results did not identify a non-conformance or a complaint out of box (coob).
 
Event or Problem Description
It was reported that patient underwent revision surgery due to implant fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMN FEMORAL NAIL, CCD 130, RIGHT, 11.5 MM, 21.5 CM
Common Device Name
ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
MDR Report Key12237117
Report Number0009613350-2021-00352
Device Sequence Number17047079
Product Code HSB
UDI-Device Identifier00889024298927
UDI-Public00889024298927
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K091566
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47-2493-212-11
Device Lot Number2973566
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2021
Initial Report FDA Received Date07/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-