SMITH & NEPHEW, INC. GII CR DEEP FLX ISR TR S3-4 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
| Model Number |
71433443 |
| Medical Device Problem Codes |
Break (1069); Naturally Worn (2988); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode. the device is worn and damaged with scratches, cracks and chips on the surface, rendering it inoperable.The device was manufactured in shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions. a review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident. at this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that, one (1) g2 dish art tr sz7-8 11mm, one (1) g2 ins spac tr sz7-8 13mm, four (4) gii cr deep flx isr tr s3-4 10, three (3) gii cr deep flx isr tr s5-6 10, two (2) gii cr deep flx isr tr s7-8 10 and two (2) gii cr deep flx isr tr s7-8 12 are old and scuffed up.After visual inspection of two gii cr deep flx isr tr s3-4 10 ((b)(4) and (b)(4)) were found to be cracked, rendering them inoperable.No case involved; therefore, there was no patient involvement.
|
| |
|
Search Alerts/Recalls
|
|
|