• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US POSTERIOR LUMBAR CAGE 8MM X 23MM X6DEG - 9MM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US POSTERIOR LUMBAR CAGE 8MM X 23MM X6DEG - 9MM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number 48950086
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Hospital disposed of device.
 
Event or Problem Description
It was reported that during placement of a tritanium pl cage intra-operatively, the positioning of the implant was not optimal.The implant "pulled out to the ventral side".To remove the mispositioned cage, the patient was operated on laterally.The event resulted in a 60 minute surgical delay for the additional surgery required.
 
Event or Problem Description
It was reported that during placement of a tritanium pl cage intra-operatively, the positioning of the implant was not optimal.The implant "pulled out to the ventral side".To remove the mispositioned cage, the patient was operated on laterally.The event resulted in a 60 minute surgical delay for the additional surgery required.
 
Additional Manufacturer Narrative
Visual, dimensional, functional and material analysis could not be performed as the device was disposed by the customer.Device and complaint history records were reviewed for this lot, and no relevant manufacturing issues or similar complaints were identified.The surgical technique provides the proper steps for sizing, trailing, and inserting the implants for when distraction is released, along with the steps to remove the implant if necessary.Per additional correspondence with the field, it is unknown how the disc space was distracted/prepped, if compression was applied to disc space after cage insertion, or if trailing was performed.Operative notes, photos and x-rays were not available.As further information about the case was not provided, a definite cause cannot be determined.Possible causes include excessive tamping during insertion, improper sizing/trailing/distraction, and/or poor patient bone quality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSTERIOR LUMBAR CAGE 8MM X 23MM X6DEG - 9MM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key12239868
Report Number3004024955-2021-00021
Device Sequence Number12113318
Product Code MAX
UDI-Device Identifier07613327117349
UDI-Public07613327117349
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number48950086
Device Catalogue Number48950086
Device Lot NumberE4K5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2021
Supplement Date Received by Manufacturer10/19/2021
Initial Report FDA Received Date07/28/2021
Supplement Report FDA Received Date11/16/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Patient RaceAsian
-
-