| Model Number |
TRM-240 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
05/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii, rupture, and an exchange of textured device due to patient's concern with product.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture baker grade iii, rupture, and an exchange of textured device due to patient's concern with product.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: broken shell, underweight, red particles on the surface of the shell, part of the shell is missing.A microscopic analysis was performed which identify sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Wear abrasion.Based on the device analysis the final assessment is: a sharp broken on radius assessed as unidentified (tear) opening.Missing shell assessed as inconclusive.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|