Pentax medical became aware of a report for an event which occurred in the usa stating, "no video image/error message, scope not connecting" involving pentax model ec-3890li/serial (b)(4).Additional information from the user stated the event occurred during pre-inspection check.No further information was provided.The device was returned to pentax on 07/07/2021.Pentax service inspectional findings included: image intermittent, passed dry/wet leak tests, ccd circuit board loose center connection, #3 remote control button sticking.Repairs were performed on the device which included replacement of the following components: o-rings and seals, remote control button,pcb for ccd drive pb-free, electrical connector assy.The device has been routinely serviced by pentax since install.
|