DEPUY ORTHOPAEDICS INC US PFC* FLUT TIB ROD TRL 75 X 12; KNEE INSTRUMENT : TIBIAL STEM TRIALS Back to Search Results
Model Number 86-6875
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Customer is complaining the instruments.The trials no longer get clean.Dirt still comes out of the grooves at the end of the trials even after cleaning.Seen in cssd.No patient involvement, no surgery delay.See attached photos.
 
Additional Manufacturer Narrative
Product complaint # : pc-(b)(4).Investigation summary : examination of the returned device was unable to confirm the reported event.However, the investigation found the black plastic ring component was missing.The investigation did not establish that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
KNEE INSTRUMENT : TIBIAL STEM TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12240575
Report Number1818910-2021-16344
Device Sequence Number17417653
Product Code HWT
UDI-Device Identifier10603295217572
UDI-Public10603295217572
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number86-6875
Device Catalogue Number866875
Device Lot NumberH0205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2021
Initial Date Received by Manufacturer 07/28/2021
Supplement Date Received by Manufacturer07/30/2021
Initial Report FDA Received Date07/28/2021
Supplement Report FDA Received Date07/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-