• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL CONTROL UNIT DYONICS 25; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL CONTROL UNIT DYONICS 25; ARTHROSCOPE Back to Search Results
Model Number 7211010S
Medical Device Problem Codes Increase in Pressure (1491); Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a shoulder arthroscopy, the control unit pressure increased above specifications by itself and had pressure difference readings from p2 sensor.The procedure was successfully completed without delay using the same device.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of over pressurization could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 2 hour burn-in on wet station utilizing low and high pressure and flow settings.At no time during functional testing did pump over pressurize, fluctuate, or become intermittent.All functions perform as expected.Pressure and flow readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Internal complaint reference: case-2021-00062826-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL CONTROL UNIT DYONICS 25
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key12241805
Report Number1643264-2021-02230
Device Sequence Number3962498
Product Code HRX
UDI-Device Identifier03596010555960
UDI-Public03596010555960
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
K051326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7211010S
Device Catalogue Number7211010S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2021
Initial Date Received by Manufacturer 07/06/2021
Supplement Date Received by Manufacturer08/25/2021
Initial Report FDA Received Date07/28/2021
Supplement Report FDA Received Date08/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-