| Model Number |
8932PG2 |
| Medical Device Problem Code |
Component or Accessory Incompatibility (2897)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complainant indicated that the sample is available to be returned but to date it has not been received for evaluation. if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the new substitute red sharps container in each theatre pose a safety risk.The lids do not properly fit on the red buckets and scrub nurses are unable to safely dispose of their sharps using the foot pedal as it does not lift the lid.Also there is no opening/hole in the lid to place small sharps.The end user must use their hands to lift the lid and risks a needle stick/sharp injury.In addition, they are unable to see how full the bin is prior to opening.The customer feels that this is not an appropriate safe substitute.There have been no injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Section g3 (date received by manufacturer) has been corrected from 26-jul-2021 to 22-jul-2021.
|
| |
|