| Model Number |
375-545-000 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to stryker sustainability solutions for evaluation.The device investigation is currently underway.The results will be provided in a supplemental mdr.
|
| |
|
Event or Problem Description
|
|
It was reported during a knee arthroscopy the tomcat cutter tip broke off in the patient.The surgeon was successfully able to retrieve the broken piece out of the knee cavity using metal forceps.There was no patient injury or medical intervention and extended procedure time reported was less than 5 minutes.The broken piece was retrieved.These are commonly used devices that are readily available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device it was found to have two reprocessing marks present on both the inner and outer shafts.No damage to the hubs was observed.The inner shaft tip was completely detached from the shaft and not returned with the device.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: - user applying too much force to the device - the arthroscopic shaver may have come into contact with staples, clips, or another metal object, resulting in damage to the blade.The instructions for use (ifu) state: before beginning the procedure, verify compatibility of all instruments and accessories.Plug in and set up the generator according to the instructions in the manufacturers¿ manual.Select an arthroscopic shaver with size, blade, and function most appropriate for the procedure.Inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.Follow a suitable surgery protocol.Careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force.Do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulates.Do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury.The tip of the bur or cutter must be irrigated periodically (general recommendation: once a minute) to cool the blade and prevent excised tissues from accumulating.Do not run the instrument without appropriate suction for the duration of the process.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported during a knee arthroscopy the tomcat cutter tip broke off in the patient.The surgeon was successfully able to retrieve the broken piece out of the knee cavity using metal forceps.There was no patient injury or medical intervention and extended procedure time reported was less than 5 minutes.The broken piece was retrieved.These are commonly used devices that are readily available.
|
| |
|