BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. INTIMA-II Y 24GAX0.75IN PRN/EC SLM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383033
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/09/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: in response to the event reported by your facility a device history review was conducted for lot number 0322636.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.Bd will continue to monitor this issue.
 
Event Description
It was reported that intima-ii y 24gax0.75in prn/ec slm had a loose tubing clamp.The following information was provided by the initial reporter: "the nurse of the hospital was ready to place a closed vein indwelling needle for the patient.A closed precision indwelling needle was taken out of the box and the package was checked to be complete without damage.After unpacking, the parts were checked and the sealing clamp was found to be a little loose".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTIMA-II Y 24GAX0.75IN PRN/EC SLM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH 
Manufacturer (Section G)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH  
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12242745
MDR Text Key264054060
Report Number3006948883-2021-00835
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/14/2023
Device Catalogue Number383033
Device Lot Number0322636
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/29/2021
Initial Date FDA Received07/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-