• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 420-159
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Unintended Radiation Exposure (4565)
Date of Event 07/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient's date of birth unavailable.Patient''s weight unavailable.Patient's ethnicity/race unavailable.Relevant tests/laboratory data unavailable.Other relevant history unavailable.The manufacturer is anticipating the return of the device for evaluation, but has not received it at this time.
 
Event or Problem Description
A peripheral atherectomy procedure commenced to treat a severely calcified lesion in the patient's proximal superficial femoral artery (sfa).The physician chose to use a spectranetics turbo elite laser atherectomy catheter to treat the patient.After the physician had used the turbo elite device in the patient and was removing the device, he noticed the "inside components of the catheter were coming out of the catheter".A balloon was used to finish the procedure, and there was no reported patient harm.This event is being submitted to capture the turbo elite which was found damaged after use and resulted in unintended radiation exposure with potential for harm.
 
Additional Manufacturer Narrative
D9): the device was returned to the manufacturer on 10 aug 2021.G3): the device was evaluated 12 aug 2021.H3): device evaluation: the device was returned and evaluated by a cross functional team.An obvious breach to the outer jacket was seen 3cm from the device''s distal tip and extended to the guide wire port.The device''s inner lumen and fibers were exposed but no dead fibers were seen.Using an 0.14 guide wire threaded through the distal end of the device, the guide wire exited out of the inner lumen in the area of the breach.Likely, during removal of the device, the guide wire was pulled laterally away from the device which caused the outer jacket to tear.This failure mode has been determined to be use related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
MDR Report Key12242946
Report Number1721279-2021-00143
Device Sequence Number14372992
Product Code MCW
UDI-Device Identifier00813132024765
UDI-Public(01)00813132024765(17)230422(0)FBH21D12A
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K170059
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/22/2023
Device Model Number420-159
Device Catalogue Number420-159
Device Lot NumberFBH21D12A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/08/2021
Supplement Date Received by Manufacturer08/12/2021
Initial Report FDA Received Date07/28/2021
Supplement Report FDA Received Date08/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BALLOON MANUFACTURER UNK.; COOK MEDICAL 6F INTRODUCER SHEATH.; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM.; TERUMO RUNTHROUGH 0.014 GUIDE WIRE.
Patient Age85 YR
-
-