• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO BONE PIN 4.0MMX152MM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO BONE PIN 4.0MMX152MM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB00031
Medical Device Problem Codes Defective Device (2588); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set up for a navio assisted surgery, the navio bone pin 4.0mmx152mm were damaged.The procedure was completed, without delay, using the bone pins from the other set.Patient was not harmed as consequence of this problem.
 
Additional Manufacturer Narrative
The navio bone pin, part rob00031 used for treatment was not made available to the designated complaint unit for evaluation thus, a visual and functional evaluation could not be performed and a relationship between the reported event and the device could not be determined.While all products meet required manufacturing specifications prior to release a serial number or lot number is required to link the device to a dhr or nc investigation.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, factors that may have contributed to the reported event may have been associated with natural wear and tear or damage.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Event or Problem Description
It was reported that during set up for a navio assisted surgery, the navio bone pin 4.0mmx152mm was bent.The procedure was completed, without delay, using the bone pins from the other set.Patient was not harmed as consequence of this problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO BONE PIN 4.0MMX152MM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key12243538
Report Number3010266064-2021-00572
Device Sequence Number5128012
Product Code OLO
UDI-Device Identifier00885556685310
UDI-Public00885556685310
Combination Product (Y/N)N
Initial Reporter CountryIN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB00031
Device Catalogue NumberROB00031
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/14/2021
Supplement Date Received by Manufacturer10/11/2021
Initial Report FDA Received Date07/28/2021
Supplement Report FDA Received Date10/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-