Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS PROLACTIN ASSAY; RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN) Back to Search Results
Catalog Number 07027737190
Medical Device Problem Codes High Test Results (2457); Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Both the customer and investigation results were above the reference range leading to the same clinical interpretation.Product labeling states: "in case of implausible high prolactin values a precipitation by polyethylene glycol (peg) is recommended in order to estimate the amount of the biological active monomeric prolactin." peg treatment was not performed.A general reagent issue can be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event or Problem Description
The initial reporter questioned a result for 1 patient sample tested for elecsys prolactin ii (prolactin ii) on a cobas e801 module.The initial result from the customer¿s e801 module was 44 ng/ml.This result was reported outside of the laboratory where the physician requested additional testing.The sample was submitted for investigation where the result from an e801 module used at the investigation site was 57.8 ng/ml.The e801 module serial number used at the customer site is not known.The e801 module used at the investigation site was (b)(4).The prolactin ii reagent lot used at the investigation site was 482648 with an expiration date of nov-2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS PROLACTIN ASSAY
Common Device Name
RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key12244543
Report Number1823260-2021-02207
Device Sequence Number19611193
Product Code CFT
UDI-Device Identifier04015630940097
UDI-Public04015630940097
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K964748
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07027737190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2021
Initial Report FDA Received Date07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-