| Model Number |
INT585 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
05/20/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).
|
| |
|
Event or Problem Description
|
|
Doctor reported the implant after 5 months of definitive loading relocate into sinus.Doctor removed the implant opening a lateral window of the maxilar sinus.Patient felt pain.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.One osseotite® tapered certain® implant 5 x 8.5mm (int585) was returned for investigation.Visual evaluation of the as returned product identified some signs of use and debris attached to the external threads but no apparent malfunction.The device could not be functionally tested for the reported failures (perforated sinus), however, measurements were taken using a caliper.Device history record (dhr) review was completed for the subject lot number 2017020656.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot (2017020656) using keyword relocation into sinus, perforated sinus and no other complaint was found.Therefore, based on the available information and dimensional analysis, device malfunction did not occur.Additionally, the reported events could not be verified since they are medical conditions.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|