• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH DRILL SLEEVE, LOCKING 3.1MM, SHORT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH DRILL SLEEVE, LOCKING 3.1MM, SHORT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 705004
Medical Device Problem Codes Separation Failure (2547); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: "the drill and sleeve stuck during drilling".
 
Event or Problem Description
As reported: "the drill and sleeve stuck during drilling.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the reported failure mode.The device inspection revealed the following: the drill and the drill sleeve were found to be stuck with each other.After dis-assembling, the drill sleeve showed normal signs of usage.Proximal opening of the sleeve appears degraded, most likely due to interaction with the drill.Minimal clearance and misalignment of the drill may have led to such a situation.A combination of friction and metal debris generated potentially clogged the assembly.A review of the device history for the reported lot did not indicate any abnormalities.A review of the labeling did not indicate any abnormalities.Based on the investigation, with the extent of damage, it can be ascertained that the root cause of the issue is user related.Drill getting stuck inside the sleeve is a direct result of excessive friction between the drill and the sleeve due to misalignment during use.If any further information is provided, the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL SLEEVE, LOCKING 3.1MM, SHORT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key12246318
Report Number0008031020-2021-00321
Device Sequence Number4600315
Product Code LXH
UDI-Device Identifier07613327083606
UDI-Public07613327083606
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number705004
Device Catalogue Number705004
Device Lot NumberK0A0748
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2021
Initial Date Received by Manufacturer 07/07/2021
Supplement Date Received by Manufacturer09/29/2021
Initial Report FDA Received Date07/29/2021
Supplement Report FDA Received Date09/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-