MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60WF
Medical Device Problem Code Material Opacification (1426)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.There are two medical device reports associated with this patient.This report is associated with the left eye.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that following an intraocular lens (iol) implant procedure, there was iol opacification.The surgeon plans to exchange the iol.Additional information has been requested; however, further information has not been received.
 
Additional Manufacturer Narrative
The product was not returned for analysis.The 2 slit lamp photos are grainy in appearance with low resolution.One right eye (od) and one left eye (os) image are very similar presentations however, od photo has a much better quality and focus (os image seems blurred or out of focus).Both slit lamp photos are in direct lighting, with a large conical beam focused on what appears to be the anterior surface of the iol (intraocular lens).There are large refractile particles on both images od>os.Due to the lighting, it is difficult to determine the location (anterior surface, posterior surface, or within the iol) of these particles.Particles appear to be larger in size and more diffuse in presentation when compared with microvacuoles (typically located within the body of the iol).Particles of this size are typically located on the front and/or back surface of the iol and are due to proteins adhered to the iol.They appear to be more inflammatory in nature but difficult to determine the etiology of these particles.The root cause for the reported complaint could not be determined.All product and batch history records are quality reviewed prior to product release.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
MDR Report Key12247552
Report Number9612169-2021-00160
Device Sequence Number14146286
Product Code HQL
UDI-Device Identifier00380652332699
UDI-Public00380652332699
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.215
Device Lot Number21270992
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/08/2021
Supplement Date Received by Manufacturer10/05/2021
Initial Report FDA Received Date07/29/2021
Supplement Report FDA Received Date10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-