Model Number 83779
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter facility name: (b)(6).
 
Event or Problem Description
It was reported that a foreign metal was detached from the protective sheath.A 3mmx6cm interlock coil was selected for use.During the procedure, embolization was performed with a total of three free coils and one controllable coil were used.Consequently, after opening the package of the 3mmx6cm coil, it was found that a foreign metal detached from the protective sheath.The procedure was completed with another of same device.There were no patient complications reported and the patient was stable.
 
Additional Manufacturer Narrative
E1.Initial reporter facility name: (b)(6) college.Device evaluated by manufacturer: the device was returned for analysis.A pusher wire and introducer sheath were returned for this complaint.The twist lock of the introducer sheath had been opened.The introducer sheath was inspected, and no anomalies were noted.The pusher wire was inspected, and no damages was found.The proximal end has a smooth surface.The interlocking arm was inspected, and no anomalies were noticed.
 
Event or Problem Description
It was reported that a foreign metal was detached from the protective sheath.A 3mmx6cm interlock coil was selected for use.During the procedure, embolization was performed with a total of three free coils and one controllable coil were used.Consequently, after opening the package of the 3mmx6cm coil, it was found that a foreign metal detached from the protective sheath.The procedure was completed with another of same device.There were no patient complications reported and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12247569
Report Number2134265-2021-09753
Device Sequence Number2796132
Product Code KRD
UDI-Device Identifier08714729764939
UDI-Public08714729764939
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2023
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0026442887
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2021
Initial Date Received by Manufacturer 07/12/2021
Supplement Date Received by Manufacturer08/18/2021
Initial Report FDA Received Date07/29/2021
Supplement Report FDA Received Date09/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age23 YR
-
-