| Model Number |
83779 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter facility name: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign metal was detached from the protective sheath.A 3mmx6cm interlock coil was selected for use.During the procedure, embolization was performed with a total of three free coils and one controllable coil were used.Consequently, after opening the package of the 3mmx6cm coil, it was found that a foreign metal detached from the protective sheath.The procedure was completed with another of same device.There were no patient complications reported and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter facility name: (b)(6) college.Device evaluated by manufacturer: the device was returned for analysis.A pusher wire and introducer sheath were returned for this complaint.The twist lock of the introducer sheath had been opened.The introducer sheath was inspected, and no anomalies were noted.The pusher wire was inspected, and no damages was found.The proximal end has a smooth surface.The interlocking arm was inspected, and no anomalies were noticed.
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign metal was detached from the protective sheath.A 3mmx6cm interlock coil was selected for use.During the procedure, embolization was performed with a total of three free coils and one controllable coil were used.Consequently, after opening the package of the 3mmx6cm coil, it was found that a foreign metal detached from the protective sheath.The procedure was completed with another of same device.There were no patient complications reported and the patient was stable.
|
| |
|