| Model Number |
7000TFX |
| Medical Device Problem Codes |
Gradient Increase (1270); Perivalvular Leak (1457)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/06/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, the nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.Multiple requests for additional details regarding this event (i.E., medical records) have been performed; however, no additional information has been provided.The subject device is also unavailable for evaluation as it remains implanted in the patient.Therefore, the reported event could not be confirmed.There is currently insufficient information to determine the root cause of this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.If any additional information is received, a supplemental report will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 7000tfx pericardial valve, implanted in the mitral position, underwent a valve-in-valve procedure after an implant duration of 12 years, three (3) months due to stenosis and paravalvular leak.The tmvr was performed successfully with a 26mm 9750tfx transcatheter valve.No additional information has been provided.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm edwards pericardial valve, implanted in the mitral position, underwent a valve-in-valve procedure after an unknown implant duration due to stenosis and paravalvular leak.The tmvr was performed successfully with a 26mm 9750tfx transcatheter valve.No additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturer narrative: the device history record (dhr) could not be reviewed, as the device serial number was not provided.Updated h6 device code(s) by adding code-1457.Corrected h6 clinical code by replacing code-4581 with code-4580.Corrected h6 investigation conclusions by removing code-22.
|
| |
|