| Model Number |
SK700 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting use to the microsnare during procedure to treat little calcified plaque in the distal common iliac artery with chronic total occlusion (100%).The vessel was moderately tortuous with diameter 8mm and length 40mm.A non-medtronic (terumo stiff angled glidewire) guidewire was used in the procedure.The device was inspected and prepped as per ifu with no issues noted.It was reported that the snare hoop broke in half while trying to capture the wire.The broken snare and wire were removed from the patient through the sheath.There was still a thread of wire in the sheath and that was removed when the sheath was pulled manually and pressure held in post op.There was no vessel damage noted.No intervention required for the removal of the snare from the patient.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation visual inspection: biologics were observed on the device.The snare returned with one end of the hoop detached from the snare.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|