RESPIRONICS, INC BIPAP AUTOSV ASSY 60 SRS 30CM,DOM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
|
Back to Search Results |
|
| Model Number |
960T |
| Medical Device Problem Codes |
Contamination (1120); Degraded (1153)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Headache (1880); Inflammation (1932); Dizziness (2194); Dry Mouth (4485); Unspecified Kidney or Urinary Problem (4503); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a bilevel positive airway pressure (bipap) device's sound abatement foam.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported an allegation of an issue related to sound abatement foam.This action was reported to the fda per 21 cfr part 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected section b5 should be reported as: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging black particles in his device, headaches, diffculty taking a breath,ocassional dry throat related to a bipap device's sound abatement foam.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting an updated report at this time.If pertinent information becomes available to the manufacturer at a later date, a follow-up report will be filed.Section h6 clinical coding corrected and updated in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received on oct 09, 2023, and section h11 should be reported as: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging an issue related to a cpap device's sound abatement foam.The manufacturer received information alleging blackparticles in his device, headaches, diffculty taking a breath,ocassional dry throat related to a bipap device's sound abatement foam.Additonal informatin was received bout the allegation of dizziness, inflammatory response and kidney disease or toxicity.The section e1 was updated and sections b1, b2, h1, h6 have been corrected in this report as follows: section b1 was corrected for adverse event and product problem.(only the product problem was checked in the previous mdr.) section b2 was corrected to other serious or important medical events.(previously, it was blank.) section h1 was changed from malfunction to serious injury.Section h6 health effects: clinical codes and health effects - impact code was corrected.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging an issue related to a cpap device's sound abatement foam.The manufacturer received information alleging black particles in his device, headaches, difficulty taking a breath, ocassional dry throat related to a bipap device's sound abatement foam.There was no report of patient harm or injury.During investigation the manufacturer observed an unknown dust contaminant on the top enclosure, pca, side panel, and blower.Hair-like particles were observed on the side panel and blower.The blower seal was observed to have a clear sticky residue on it, as well as a hair-like particles.The blower impellers were observed to have an unknown dust contaminant on them inside the blower, confirming debris in the airpath.An hair-like particles were also observed on the right-side panel, bottom enclosure, alarm buzzer, and air inlet seal.The sound abatement foam was confirmed to have degraded and was attached to the underside of the lower blower housing and on the bottom enclosure.The manufacturer used a fitt (foam integrity test tool) and was able to confirm the presence of degraded sound abatement foam.The device's event logs were downloaded and reviewed by manufacturer.The manufacturer found no error codes.The manufacturer applied power to the device and verified airflow.The manufacturer concludes, there was evidence of sound abatement foam degradation/breakdown and observed multiple contaminants and was able to confirm the complaint or address the symptoms specified.Sections d8, d9, h2, h3, h6 have been updated in this report.
|
| |
|