The reported event was inconclusive.No sample was returned for evaluation.Potential root cause for this failure mode could be user related (example: salt accumulation)/block drainage lumen/no drainage eye).The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿to deflate catheter balloon.Gently, insert a syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If the tails, contact an adequately trained professional for assistance as detected by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.¿.
|