• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 5/200/130; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 5/200/130; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 376301
Medical Device Problem Code Decrease in Pressure (1490)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A passeo-18 balloon catheter was chosen for treatment.It was attempted to inflate the balloon up to 5 bar, but to keep it at 5 bar it was necessary to give more pressure.Eventually the passeo-18 had to be replaced by another balloon as it could not be used anymore.Lot number and event date are unavailable.
 
Additional Manufacturer Narrative
Only the material number was reported whereas the affected instrument was discarded.Even after additional correspondence with the local staff, neither a lot number nor an event date could be obtained.An investigation with regards to the reported complaint event could therefore not be performed.During the manufacturing process, every lot is tested by predefined specifications in the in-process and final inspections.During final inspection, each instrument is tested for air tightness by means of a helium leak test and a pressure test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 5/200/130
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key12252899
Report Number1028232-2021-04308
Device Sequence Number5244731
Product Code LIT
UDI-Device Identifier07640130430082
UDI-Public07640130430082
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number376301
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/29/2021
Supplement Date Received by Manufacturer09/03/2021
Initial Report FDA Received Date07/30/2021
Supplement Report FDA Received Date09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-