| Model Number |
376301 |
| Medical Device Problem Code |
Decrease in Pressure (1490)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A passeo-18 balloon catheter was chosen for treatment.It was attempted to inflate the balloon up to 5 bar, but to keep it at 5 bar it was necessary to give more pressure.Eventually the passeo-18 had to be replaced by another balloon as it could not be used anymore.Lot number and event date are unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
Only the material number was reported whereas the affected instrument was discarded.Even after additional correspondence with the local staff, neither a lot number nor an event date could be obtained.An investigation with regards to the reported complaint event could therefore not be performed.During the manufacturing process, every lot is tested by predefined specifications in the in-process and final inspections.During final inspection, each instrument is tested for air tightness by means of a helium leak test and a pressure test.
|
| |
|