Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the end of the cable is stripped.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: cable sheath cut.Corroded motor.Keyboard resistance value out of tolerance limits.The motor, motor cable and keyboard were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
It was reported by the affiliate in that during an unknown surgery on an unknown date, it was observed that the device was.During in-house engineering evaluation, it was determined that the motor was corroded on the device.There was no procedure nor patient involvement reported.No additional information was provided.
|