DEPUY SPINE INC VIPER PRIME CFXFN XTB 6X40MM S; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
|
Back to Search Results |
|
| Model Number |
186760440S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional narrative: additional device product codes: nkb, kwp.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.(b)(4).Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that the patient underwent a revision procedure on (b)(6) 2021 as two (2) implanted viper prime screws broke in s1 level postoperatively.Hardware were explanted.Patient outcome is unknown.No further information provided.This report is for one (1) viper prime cfx fen x-tab polyaxial screw 5.5 6 x 40mm.This is report 2 of 2 for complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
Concomitant devices: unknown rod (part# unknown, lot# unknown, qty 2).Unknown locking/set screw (part# unknown, lot# unknown, qty unknown).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Photo investigation: the device was not returned.A photo-investigation was performed on the x-ray images.Upon inspecting the x-ray images provided, the viper prime screws were observed to have been implanted.However, the reported complaint condition of broken screws cannot be confirmed from the provided images.The root cause for the reported event cannot be determined from the available information.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Conclusion: the complaint condition cannot be confirmed during photo investigation.During the investigation, no product design issues or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|