• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYMMETRY SURGICAL STERILIZATION TRAY, SINGLE LAYER CYSTOSCOPY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYMMETRY SURGICAL STERILIZATION TRAY, SINGLE LAYER CYSTOSCOPY Back to Search Results
Model Number ST-CR
Medical Device Problem Codes Break (1069); Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was received at olympus (b)(6) site for initial evaluation and investigation.A visual inspection on the as is received condition of the device was performed.The device was returned in a brown box which has several creases throughout the box.Inside the box contained some bubble wrap and thin padded sheets that covered the tray.The tray appears to have no scratches, dents or deformities however, it was observed that one side of the tray where the handles are located, the metal rod that holds the clip in place is out of place and not securely locked in.It was verified that the metal rod is missing a washer on each side to keep the metal rod securely in position.The missing washers were not returned for evaluation and was not found inside the brown box or tray which returned with the device.It was confirmed that the other handle on the other side of the tray has a washer on both ends of the metal rod and is securely locked in place.The subject device is being shipped to the oem (original equipment manufacturer) for investigation.This report will be supplemented accordingly following investigations.
 
Event or Problem Description
It was reported that during installation, the handle was broken on the lid.There was no patient involvement, no user injury reported due to the event.This report is being submitted for the handle was broken on the lid, metal rod is missing a washer on each side to keep the metal rod securely in position.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The subject device was returned to the oem (original equipment manufacturer) for evaluation.Investigation from oem below: the critical step was accomplished by an experienced operator and followed the approved work instruction.The lid was an injected molded part which requires less preparation for proper assembly and less applicable failure modes than if it were a thermoformed part.Reviewed returned components without no objective evidence of sterilization.Focused on the lid for any evidence of form fit or function and also the pin and assembly.The most likely assignable cause is that the retention washers were never assembled with this one injection molded lid.The dhr (device history record) reconciliation of components indicated that the washers were available for assembly and that no additional parts were left over.This product is sampled at final and the sample inspection did not discover the defect issue.The dhr was reviewed for the product and there was nothing on the dhr review that would lead to the reported issue.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILIZATION TRAY, SINGLE LAYER CYSTOSCOPY
Common Device Name
STERILIZATION TRAY, SINGLE LAYER CYSTOSCOPY
Manufacturer (Section D)
SYMMETRY SURGICAL
253 abby road
manchester NH 03103
MDR Report Key12254082
Report Number3003790304-2021-00129
Device Sequence Number3681264
Product Code FSM
UDI-Device Identifier00821925007857
UDI-Public00821925007857
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K944025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberST-CR
Device Lot Number000216591
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/08/2021
Supplement Date Received by Manufacturer08/03/2021
Initial Report FDA Received Date07/30/2021
Supplement Report FDA Received Date08/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-