MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
APB-1.5-4-HX-ES |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding an axium coil that broke in the microcatheter.The patient was undergoing a coil embolization procedure to treat a ruptured fusiform aneurysm in the left middle cerebral artery.The aneurysm max diameter was 2.1mm and the neck diameter was 1.7mm.Blood flow was normal.Vessel tortuosity was moderate.It was reported that the axium coil and all accessory devices were prepared as indicated in the instructions for use (ifu).The catheter was flushed continuously.During delivery of the coil, there was friction and the coil was detached and an damaged.It was noted that the physician did not reposition the coil during the procedure.The coil was removed and replaced to continue the procedure.There were no patient symptoms or complications associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The coil was detached in the microcatheter.The catheter was echelon10, lot# unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the axium pushwire was returned for evaluation.There was no implant coil or catheter returned with the pushwire.The implant coil appeared to be detached from the pushwire.The shield coil was present and intact.The coin was not present against the lumen stop as it was pulling back.The pusher was found broken at the distal break indicator (manual detachment location); with the release wire pulling back; it appeared that the manual detachment was attempted at this location.The break indicator, ai and coupler tubing were found missing and not returned.Under the microscope, the outer jacket was then removed to gain access the coin.The coin was measured in 3 locations (0.077mm @ 0.063mm; measured 0.089mm @ 0.127mm; measured 0.099mm @ 0.275mm) and found to be within specifications.The inner diameter of the lumen stop, and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00275¿ and found to be within specification.The retainer ring inner diameter (id) was measured to be 0.00460¿ and found to be within specification.All other subassemblies appeared to be normal.No other anomalies were observed.Based on the analysis performed, the axium prime coil was not confirmed to have "coil kink/damage" and "coil resistance in catheter" as the pushwire was returned without the coil.However, the root cause could not be determined.Possible causes include damaged catheter, vessel tortuosity and lack of continuous flush with heparinize saline during delivery.Regarding to the "premature detachment" issue, the customer complaint was confirmed as the pushwire was returned with the implant coil already detached.The root cause could not be determined.Based on the returned device, there was evidence of manual detachment attempt to detach the coil as the pusher was also found to be broken at the manual detachment location; with the coin pulled back from the lumen stop.The pulling back of the coin from the lumen stop may have contributed to the detachment of the implant coil from the pushwire.There was no non-conformance to specification identified that that led to the premature detachment issue.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Since the implant coil, break indicator, ai and coupler tubing were not returned; any contribution of the implant coil, break indicator, ai and coupler tubing to the premature detachment issue could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|