It was reported a quick-core coaxial biopsy needle set was required for an unknown procedure.Prior to patient contact, the stylet was screwed too tight in the outer part of the needle and could not be separated.The procedure was completed with a similar device.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation.It was reported that a quick-core coaxial biopsy needle set (qcs-18-15.0-20t) from lot 13703499 was screwed too tight in the outer part of the needle.This failure was noticed prior to patient contact.Cook became aware of this event upon being notified by nanjing changde med.Supp.Co.In china.The procedure was completed with a replacement device.The patient did not experience any adverse effects.A review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures of the device, as well as a visual inspection and functional test of the returned product, were conducted during the investigation.One prior to use qcs set was returned.The trocar was unable to be unscrewed.Additionally, a document based investigation evaluation was performed.Review of the device master record (dmr) concluded there are appropriate controls in place to detect damage prior to distribution.A review of the device history record (dhr) was also conducted to check for failure related nonconformances and additional complaints.There are no related nonconformances or additional complaints.The device is shipped with instruction for use (ifu) which provides the following information to the user related to the reported failure mode: "how supplied supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.".Based on the device history record, device master record, device failure analysis, and product labeling review, there is no indication the complaint device was manufactured out of specification.There is no evidence of nonconforming product in house or in the field.It is possible the device was screwed too tightly during quality control, but this potential cause of failure cannot be confirmed without additional information.There is no evidence of manufacturing deficiency.Based on the information provided, the examination of returned product and the results of the investigation, the cause was traced to a component failure.The appropriate internal personnel have been notified.Per the risk assessment, no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|