| Model Number |
IPN913839 |
| Medical Device Problem Codes |
Fracture (1260); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported "when they use the circuits with this lot number, the vent throws an alarm until the circuit is changed to one not from this lot number.Pt care is not affected except by the alarm noise.No clinical deterioration".Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer returned one niv circuit part number 870-98kitf (22mm heated niv circ kit w/filter) for analysis.While performing the visual inspection test on the received sample a crack was found on the connector.Leak testing was performed and the sample failed due to the cracked connector.No other issues were found.The complaint was confirmed; however, a root cause for the issue could not be identified.Corrected data: section f.10.- medical device problem code corrected to 1260 section h.6.- medical device problem code corrected to 1260.
|
| |
|
Event or Problem Description
|
|
It was reported "when they use the circuits with this lot number, the vent throws an alarm until the circuit is changed to one not from this lot number.Pt care is not affected except by the alarm noise.No clinical deterioration".Patient condition reported as "fine".
|
| |
|