| Model Number |
10604 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.50 x 24 synergy drug-eluting stent was advanced but failed to cross the lesion and it was noted that the stent was too damaged.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy ii us mr 3.50 x 24 mm stent delivery system was returned for analysis.An examination of the crimped stent found stent damage.Stent struts from the distal stent region were noted to be lifted from their crimped position and pulled proximally.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks along the length of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.50 x 24 synergy drug-eluting stent was advanced but failed to cross the lesion and it was noted that the stent was too damaged.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|