According to the reporter, prior to dialysis treatment, a hairline cracks and leak were observed on the arterial (luer adapter) connector.Flushing was done prior to use and leakage was detected.Catheter was repaired immediately with repair kit and dialysis was made normal after repair.The repaired catheter remained in patient and only the connectors with hairline cracks will be returned.The device was used successfully after it was repaired and the procedure was completed.There was no blood loss.Octenisept alcohol free was the cleaning agent used on the device and typically utilized to clean the adapters.The cleaning agent was allowed to dry thoroughly prior to applying ointment to the area.There was no intervention/treatment required but only repair kit was used.There was no tego utilized.There was no instrument used but only hands to loosen or tighten the device.There were no other products being utilized with the device.There were no patient symptoms or complications associated with the event.There was no patient injury.
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted a cracks on the arterial luer adaptor.It was reported that the luer adapter was leaking.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|