| Model Number |
LC250 |
| Medical Device Problem Codes |
Fire (1245); Smoking (1585); Sparking (2595)
|
| Health Effect - Clinical Code |
Superficial (First Degree) Burn (2685)
|
| Date of Event |
07/31/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been made available for evaluation.Should the device or further information become available, a follow-up report will then be issued.
|
| |
|
Event or Problem Description
|
|
Alleges the hand control caught fire.Alleges she was pushing the buttons to get out of the chair when it started smoking and sparking, then caught fire.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and evaluated.Misuse-failure to periodically check hand control for damage.The root cause of the failure appears to be a damaged harness.The operation precautions in the manual instructs to periodically check the hand control and all power cords for visible damage.
|
| |
|
Event or Problem Description
|
|
Alleges the hand control caught fire.Alleges she was pushing the buttons to get out of the chair when it started smoking and sparking, then caught fire.
|
| |
|