Model Number LC250
Medical Device Problem Codes Fire (1245); Smoking (1585); Sparking (2595)
Health Effect - Clinical Code Superficial (First Degree) Burn (2685)
Date of Event 07/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not yet been made available for evaluation.Should the device or further information become available, a follow-up report will then be issued.
 
Event or Problem Description
Alleges the hand control caught fire.Alleges she was pushing the buttons to get out of the chair when it started smoking and sparking, then caught fire.
 
Additional Manufacturer Narrative
The device was returned and evaluated.Misuse-failure to periodically check hand control for damage.The root cause of the failure appears to be a damaged harness.The operation precautions in the manual instructs to periodically check the hand control and all power cords for visible damage.
 
Event or Problem Description
Alleges the hand control caught fire.Alleges she was pushing the buttons to get out of the chair when it started smoking and sparking, then caught fire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
MDR Report Key12259143
Report Number2530130-2021-00084
Device Sequence Number4848310
Product Code INO
UDI-Device Identifier00606509400237
UDI-Public00606509400237
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K070950
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberLC250
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2021
Supplement Date Received by Manufacturer08/02/2021
Initial Report FDA Received Date08/02/2021
Supplement Report FDA Received Date09/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-