• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 977A275
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that during lead placement, the stylet was removed and reinserted.The stylet would not fully advance despite the surgeon¿s careful handling and numerous efforts.Another lead was used as a result to resolve the issue.No symptoms were reported.
 
Additional Manufacturer Narrative
Product analysis #(b)(6): analysis information -- 2021-08-23 14:20:46 cst pli# 10 product id# 977a275 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis identified that there was a gap in the stylet coil of the lead.H3: analysis of the lead (s/n (b)(6)) identified that there is a gap or kink in the stylet coil 10.5cm from the distal end of the lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VECTRIS SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key12259605
Report Number2649622-2021-15433
Device Sequence Number4742981
Product Code LGW
UDI-Device Identifier00763000006525
UDI-Public00763000006525
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/08/2024
Device Model Number977A275
Device Catalogue Number977A275
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2021
Supplement Date Received by Manufacturer08/23/2021
Initial Report FDA Received Date08/02/2021
Supplement Report FDA Received Date08/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
-
-