It was reported to olympus that during reprocessing, the channel-opening cleaning brush lost bristles.When removing the brush from the packaging, the bristles were messy and a little loose.The brush was passed through the equipment, and it was destroyed.All of the bristles were stuck in the equipment.The brush was one (1) of five (5) in which the same issue occurred.Four (4) of the five (5) brushes were discarded.None of the brushes went through sterilization reprocessing, but rather at the first use, the bristles came out.The sterilization staff knew how to use the brush.It was purchased for a long time.There was no new staff or changes in the process.The intended procedure was completed.The same device was not used to complete the procedure.No other equipment was replaced.No patient harm reported.This complaint is related to patient identifiers: (b)(6) (model # mh-507, serial number: (b)(4)), (b)(6) (model # mh-507, serial number: unknown), (b)(6) (model # mh-507, serial number: unknown), (b)(6) (model # mh-507, serial number: unknown).
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation.The dhr review could not be performed as no lot number was reported and there was no device return to inspect for a lot number.If the lot number becomes available at a later date, the dhr shall be reviewed.The root cause could not be conclusively specified.It was likely the issue was caused by the bending of the shaft, which created a gap in the part where the hair was bundled.Bending of the shaft occurred when an external force was applied to the brush during the first use.Alternatively, it was likely the issue was caused by the application of external force during transportation or storage to the facility.The instructions for use (ifu) provides the following guidelines: when inserting the channel-opening cleaning brush (mh-507), the single use channel-opening cleaning brush (maj-1339), or the single use combination cleaning brush (bw-412t) into the suction cylinder, do not forcibly insert the brush beyond the middle of the brush section.Otherwise, the brush may become stuck in the suction cylinder.
|