• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL SYRINGE 20ML LL BNS; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL SYRINGE 20ML LL BNS; PISTON SYRINGE Back to Search Results
Catalog Number 301031
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/17/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: it was reported that syringes found with black and white dots and are damaged or broken.To aid in the investigation, ten samples and nine photos were received for evaluation by our quality team.The samples came bulk packaged.A visual inspection was performed and damaged syringe barrels, embedded degraded resin and rubmarks were observed.Rubmarks are imperfections and considered acceptable.The photos provided show some of the samples received.The embedded degraded resin in the component typically occurs at the startup or intermittently during the injection molding process.The degraded resin can break loose and be molded into components.For the damaged parts defect, it is likely that a jam occurred during the assembly process resulting in the damage to the syringes.As the lot provided is 'unknown,' a device history record review could not be completed.Verification of the assembly equipment was performed.Alignment of rails and conveyors were acceptable.Flow of products was good.Frequency of inspections were increased to mitigate the embedded degraded resin escapes.Based on the investigation and with the returned sample analysis the symptom reported by the customer is confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported syringe 20ml ll bns had incorrect scale markings.The following information was provided by the initial reporter: "black dots: 218 pieces damaged/broken: 102 pieces".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 20ML LL BNS
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12264124
MDR Text Key265014397
Report Number1911916-2021-00777
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K980987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number301031
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2021
Initial Date FDA Received08/02/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-