H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one vaccess pta dilatation catheter has returned for evaluation.On the visual evaluation the device appeared bloody, no kinks noted to the catheter shaft no anomalies to the luers or y-body.The returned balloon noted to have a compound rupture at distally and the distal end of the balloon was detached, and the distal marker band was also detached not returned for evaluation.Inner guidewire lumen was exposed.No other anomalies were noted.No functional evaluation performed due to the nature of the complaint.Therefore, the investigation was confirmed for the reported balloon rupture as the balloon was noted to have a compound rupture.The investigation was also confirmed for the balloon detachment as the distal end the balloon and marker band was noted to be detached from the catheter and not returned for evaluation.A definitive root cause for the balloon rupture and balloon detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 04/2024), h11: h6(result, conclusion), h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|