| Catalog Number |
03183696122 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.The event was consistent with a sample handling or maintenance issue.As the issue appeared to be resolved by using a new reagent cassette, a general issue was unlikely.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a questionable low iron gen.2 result from cobas integra 400 plus analyzer serial number (b)(4).The initial result was 1.6 mol/l.The same sample was tested in another laboratory with an abbott instrument and the result was 6.1 mol/l.After the customer replaced the reagent cassette, the same sample was repeated and the result was equal to that obtained on the abbott instrument.No specific data was provided.The questionable result was not reported outside of the laboratory.
|
| |
|