• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSURE EPICARDIAL LEAD; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSURE EPICARDIAL LEAD; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) Back to Search Results
Model Number 4965-50
Medical Device Problem Codes Failure to Capture (1081); Impedance Problem (2950)
Health Effect - Clinical Codes Bradycardia (1751); Low Blood Pressure/ Hypotension (1914)
Date of Event 03/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient experienced bradycardia and symptomatic hypotension.It was also noted that the implantable pulse generator (ipg) was pacing below the lower programmed rate and the right atrial (ra) lead exhibited a unipolar impedance warning.Both the ipg and ra lead remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was further reported that the right atrial (ra) lead exhibited no capture.The lead and ipg were reprogrammed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE EPICARDIAL LEAD
Common Device Name
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key12267433
Report Number2649622-2021-15546
Device Sequence Number4440334
Product Code NVN
UDI-Device Identifier00613994501813
UDI-Public00613994501813
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P950024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date (Section B) 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/24/2021
Device Model Number4965-50
Device Catalogue Number4965-50
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/12/2021
Supplement Date Received by Manufacturer08/03/2021
Initial Report FDA Received Date08/03/2021
Supplement Report FDA Received Date08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
W1DR01 IPG, 511212 LEAD; W1DR01 IPG, 511212 LEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age21 YR
Patient Weight74
-
-