| Model Number |
400SMTHXSFT0102 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the basilar artery using a penumbra smart coil (smart coil).During the procedure, a smart coil would not advance at all from its initial position within its introducer sheath.Therefore, the smart coil was removed.The procedure was completed using other smart coils.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.Forceful advancement against this resistance likely contributed to the kink observed near the pusher assembly mid-joint.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance, and no further issues occurred after the pusher assembly mid-joint was out of the friction lock.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|
| |
|